Analyzers for rent : Thermo Environmental Instruments  : Model-51


The Thermo Model 51 (FID) is a total hydrocarbon analyzer featuring a compact chassis with microprocessor control. This unit also contains a  RTD temperature feedback mechanism that maintains all components in contact with the sample to 200 C, Auto-ignition, Auto Ranging,  RS-232 Communication, highly  visible Vacuum Fluorescent display. The Model 51 analyzer can perform automatic calibration routines at user prescribed time intervals.  Idealy suited For  EPA Method 25A Compliance Monitoring.
Ranges: 0-10 up to 0-10,000 ppm  Auto ranging
Detection Method: Flame Ionization Detection (FID)
Power Requirement: 120 V AC   60/50HZ
Sample Rate 2.0 LPM
Fuel Requirement: 100% Hydrogen, Zero Grade (Normal), OR  60/40 FID Fuel 
(on request)
Approx.  Weight: 52 lbs.
Suitable for EPA: Method 25A
Outputs Available: 0-100mV, 0-1V, 0-5V, 0-10V
Communication 25 pin RS-232 port
Keypad 15 segment pad with audible and tactile feedback